端粒酶活性的分析检测对于癌症诊断、预后治疗等具有重要意义。该研究利用G-四链体(G4) DNA的金属配合物靶向发光探针[Ir(ppy)
2
(pip)]PF
6
,构建了灵敏检测癌细胞端粒酶活性的电化学发光(ECL)生物传感器。首先将引物DNA组装在电极表面,然后
与癌细胞提取的端粒酶共同孵育,具有活性的端粒酶在引物末端生成端粒重复序列从而形成G4结构,随后[Ir(ppy)
2
(pip)]PF
6
通过与G4靶向作用结合到延长的序列中。结果表明[Ir(ppy)
2
(pip)]PF
6
的ECL信号与端粒酶活性呈现正相关,传感器的电化学发光强度与癌细胞浓度在10 ~ 5×10
5
cell/mL呈现良好的线性关系,检出限(LOD)为3 cell/mL。实验还表明了传感器具有良好的稳定性、适用性、抗干扰性和重现性,最后将该传感器用于测定真实人类血清和尿液样本中的端粒酶活性,得到了满意的结果。
The analysis and detection of telomerase activity is important for cancer diagnosis,prognosis,and treatment. In this study,an electrochemiluminescence(ECL) biosensor was constructed for sensitive detection of telomerase activity in cancer cells using a metal complex-targeted luminescent probe[Ir(ppy)
2
(pip)]PF
6
for G-quadruplex(G4) DNA. Primer DNA was first assembled on the electrode surface and then co-incubated with telomerase extracted from cancer cells. The active telomerase generated telomeric repeats at the end of the primer to form the G4 structure,and then[Ir(ppy)
2
(pip)]PF
6
bound to the extended sequence by targeting with G4. The results showed that the ECL signal of[Ir(ppy)
2
(pip)]PF
6
showed a positive correlation with telomerase activity,and the ECL intensity of the sensor showed a good linear correlation with the concentration of cancer cells in the range of 10-5×10
5
cell/mL,with the limit of detection(LOD) of 3 cell/mL. The experiments also demonstrated that the sensor had good stability,applicability,anti-interference ablity,and reproducibility. Finally,the sensor was used for the determination of telomerase activity in real human serum and urine samples,and satisfactory results were also obtained.
端粒酶活性G-四链体金属配合物电化学发光生物传感器
telomerase activityG-quadruplexmetal complexelectrochemiluminescencebiosensor
De-Jesus B B,Blasco M A. Trends Genet.,2013,29(9):513-520.
Wang Y,Sušac L,Feigon J. Cell. Mol. Life Sci.,2019,11(12):a032383.
Abraham-Punnoose J,Ma Y,Hoque M E,Cui Y,Sasaki S,Guo A H,Nagasawa K,Mao H. Biochem.,2018,57(51):6946-6955.
Ling P,Lei J,Jia L,Ju H. Chem. Commun.,2016,52(6):1226-1229.
Greider C W,Blackburn E H. Nature,1989,337(6205):331-337.
Vukašinović A R,Kotur-Stevuljević J M,Mlakar V,Sopić M D,Cvetković Z P,Petković M R,Spasojević-Kalimanovska V V,Bogavac-Stanojević N B,Ostanek B. Scan. J. Clin. Lab. Inv.,2019,79(3):188-193.
Shan S,Li J,Han G,Zhang R,Liu Z,Zhang Z. Analyst,2022,147(11):2405-2411.
Zong S,Wang Z,Chen H,Cui Y. Small,2013,9(24):4215-4220.
Ravalli A,Voccia D,Palchetti I,Marrazza G. Biosensors-Basel,2016,6(3):39.
Guo Y,Liu S,Yang H,Wang P,Fen Q. Anal. Chim. Acta,2021,1144:68-75.
Phan A T. FEBS J.,2010,277(5):1107-1117.
Li Y,Luo S,Wang H,Lai Y,Li D,Zhang Q,Huang H,Zhang P. J. Med. Chem.,2023,66(7):4840-4848.
Zhu W,Liu S,Wang Z,Shi C,Zhang Q,Wu Z,Li G,Zhu D. Molecules,2023,28(23):7914.
Castor K J,Metera K L,Tefashe U M,Serpell C J,Mauzeroll J,Sleiman H F. Inorg. Chem.,2015,54(14):6958-6967.
Olmon E D,Hill M G,Barton J K. Inorg. Chem.,2011,50(23):12034-12044.
Wu S,Wang S,Li Z,Wu C,Ma D L,Miao X. Anal. Bioanal. Chem.,2022,414(13):3755-3763.
Sprouse S,King K A,Spellane P J,Watts R J. J. Am. Chem. Soc.,1984,106(22):6647-6653.
Lowry M S,Hudson W R,Pascal R A,Bernhard S. J. Am. Chem. Soc.,2004,126(43):14129-14135.
Haghighatbin M A,Laird S E,Hogan C F. Curr. Opin. Electroche.,2018,7:216-223.
Dong P,Zhu L,Huang J,Ren J,Lei J. Biosens. Bioelectron.,2019,138:111313.
Yi Z,Wang H B,Chen K,Gao Q,Tang H,Yu R Q,Chu X. Biosens. Bioelectron.,2014,53:310-315.
Zhang H R,Wang Y Z,Wu M S,Feng Q M,Shi H W,Chen H Y,Xu J J. Chem. Commun.,2014,50(83):12575-12577.
0
浏览量
0
下载量
0
CSCD
关联资源
相关文章
相关作者
相关机构