彭侃霖, 宋旭琴, 刘戎, 张嘉慧, 周豪, 丘镜莉, 贺利民. Determination of Main Components in Polymyxins Crude Drugs by High Performance Liquid Chromatography with Evaporative Light Scattering Detector[J]. 2020, 39(5): 657-661.
DOI:
彭侃霖, 宋旭琴, 刘戎, 张嘉慧, 周豪, 丘镜莉, 贺利民. Determination of Main Components in Polymyxins Crude Drugs by High Performance Liquid Chromatography with Evaporative Light Scattering Detector[J]. 2020, 39(5): 657-661.DOI: doi:10.3969/j.issn.1004-4957.2020.05.015.
Determination of Main Components in Polymyxins Crude Drugs by High Performance Liquid Chromatography with Evaporative Light Scattering Detector
A high performance liquid chromatography with evaporative light scattering detector was developed for the determination of main components in polymyxin B and polymyxin E crude drugs.Analytes were separated on a Phenomenex Kinetex Biphenyl column(50 mm×4.6 mm id
5 μm) by gradient elution with methanol and 20 mmol/L ammonium acetate aqueous solution containing 0.6% formic acid as mobile phases at a flow rate of 0.5 mL/min.The parameters for evaporative light scattering detector were as follows:drift tube temperature:55 ℃
carrier gas pressure:20 psi
and the gain:250.Results showed that there were good linear relationships for the analytes in polymyxin B and polymyxin E in the concentration range of 10-200 μg/mL with their detection limits of 0.8 μg/mL and 1.2 μg/mL
and their quantitation limits of 2.6 μg/mL and 4.0 μg/mL
respectively.The intra batch recoveries for polymyxin B and polymyxin E were in the ranges of 100%-104% and 101%-103%
with intra batch relative standard deviations of 0.90%-2.4% and 0.90%-2.6%
respectively.The inter batch recoveries were in the ranges of 101%-103% and 101%-102%
with inter-batch relative standard deviations of 1.1%-1.9% and 1.3%-1.7%
respectively.Compared with microbiological detection method
the developed method is simpler
more precise and more accurate in the detection of analytes in crude drugs
and could be applied to the quality control of polymyxin B and polymyxin E crude drugs.
Analysis of Microplastic Release Behavior During Aging-Degradation of PBAT Based on Multi-technology Combined Detection
Progress in Analytical Methods for Phthalate Plasticizers in (Micro)plastic Release Systems and Environmental Media
Detection of Micro/Nanoplastics in Tea Bag Leachate Using Cloud Point Extraction Combined with Matrix-assisted Laser Desorption Ionization Time-of-flight Mass Spectrometry
Comparative Study of Py/GC-MS and TD/GC-MS for Determining Microplastic Mass Abundance in Soils Surrounding Waste Incineration Plants
Mechanisms,Effects,and Analytical Characterization Bottlenecks of pH-Driven Microplastic Aging in Aquatic Environments