Simultaneous Determination of Polybrominated Diphenyl Ethers and Novel Brominated Flame Retardants in Serum by Atmospheric Pressure Gas Chromatography-Tandem Mass Spectrometry
Scientific Papers|更新时间:2024-08-08
|
Simultaneous Determination of Polybrominated Diphenyl Ethers and Novel Brominated Flame Retardants in Serum by Atmospheric Pressure Gas Chromatography-Tandem Mass Spectrometry
Journal of Instrumental AnalysisVol. 43, Issue 8, Pages: 1273-1279(2024)
JIANG You-sheng,ZHU Lu,PENG Jin-ling,SONG Jia-yi,LIU Yuan,LIN Xiao-shi,KONG Ling-quan,ZHUO Lan,ZHANG Jian-qing.Simultaneous Determination of Polybrominated Diphenyl Ethers and Novel Brominated Flame Retardants in Serum by Atmospheric Pressure Gas Chromatography-Tandem Mass Spectrometry[J].Journal of Instrumental Analysis,2024,43(08):1273-1279.
JIANG You-sheng,ZHU Lu,PENG Jin-ling,SONG Jia-yi,LIU Yuan,LIN Xiao-shi,KONG Ling-quan,ZHUO Lan,ZHANG Jian-qing.Simultaneous Determination of Polybrominated Diphenyl Ethers and Novel Brominated Flame Retardants in Serum by Atmospheric Pressure Gas Chromatography-Tandem Mass Spectrometry[J].Journal of Instrumental Analysis,2024,43(08):1273-1279. DOI: 10.12452/j.fxcsxb.24052985.
Simultaneous Determination of Polybrominated Diphenyl Ethers and Novel Brominated Flame Retardants in Serum by Atmospheric Pressure Gas Chromatography-Tandem Mass Spectrometry
An atmospheric pressure gas chromatography-tandem mass spectrometry(APGC-MS/MS) method was established to simultaneously determine 8 types of polybrominated diphenyl ethers(PBDEs) and 9 types of novel brominated flame retardants(NBFRs). Serum sample of 0.3 g was pretreated with HLB column solid-phase extraction and acidic silica gel column purification,and separated by DB-5HT capillary column(15 m×0.25 mm×0.10 μm). PBDEs and NBFRs was quantitatively analyzed by APGC-MS/MS. The detection of PBDEs is using isotope dilution method,and the detection of NBFRs is using the internal standard method. This method was further validated using international serum standard reference material sample SRM 1958. Compared with the reference mass fraction range provided by SRM 1958,the measured concentrations of PBDEs were within the reference mass fraction range,with a relative standard deviation(RSD,
n
=6) of 6.1% to 18%. The low,medium,and high concentrations of NBFRs were added to the matrix for validation. The recoveries of matrix spiking ranged from 81.7% to 126%. The RSDs were between 6.2% and 18%. This method has high sensitivity and accuracy,stable performance,which meets the requirements for the determination of PBDEs and NBFRs in human serum and the method can be applied to human health risk assessment for PBDEs and NBFRs.
关键词
多溴二苯醚新型溴代阻燃剂血清大气压气相色谱-串联质谱
Keywords
polybrominated diphenyl ethersnovel brominated flame retardantsserumatmospheric pressure gas chromatography-tandem mass spectrometry
references
Salamova A,Hermanson M H,Hites R A. Environ. Sci. Technol.,2014,48(11):6133-6140.
Wang Y,Sun Y,Chen T,Shi Z X,Zhou X Q,Sun Z W,Zhang L,Li J G. J. Chromatogr. B,2018,1099:64-72.
Yang Y,Xie Q,Liu X. Ecotoxicol. Environ. Saf.,2015,115 :55-61.
Li X Y,Liu J H,Zhou G Q,Sang Y J,Zhang Y,Li J,Shi Z X,Zhou X Q,Sun Z W. Environ. Int.,2021,146:106307.
Sun Y M,Wang Y W,Liang B L. Sci. Total Environ.,2020,705:135783.
Jing L,Sun Y M,Wang Y W,Liang B L,Chen T,Zheng D,Zhao X Z,Zhou X Q,Sun Z W,Shi Z X. Chemosphere,2019,223:675-685.
Papachlimitzou A,Barber L J,Losada S,Bersuder J R. J. Chromatogr. A,2012,1219:15-28.
Shi Z X,Zhang L,Li J G,Wu Y N. Chemosphere,2018,198:522-536.
Li J,Chen T, Wang Y W,Shi Z X,Zhou X Q,Sun Z W,Wang D J, Wu Y N. J. Sep. Sci.,2017,40(3):709-716.
Wang Q H,Yuan H D,Jin J,Li P,Ma Y L,Wang Y. Environ. Sci.(王庆华,袁浩东,金军,李鹏,马玉龙,王英. 环境科学),2018,39(3):1438-1444.
Method EPA 1614A. Brominated Diphenyl Ethers in Water,Soil,Sediment,and Tissue by HRGC/HRMS.United States Environmental Protection Agency Standards
HRGC/HRMS检测水中、土壤、沉积物和组织中的多溴二苯醚. 美国国家环境保护局标准.
Gustavsson J, Nguyen M A,Josefsson S,Wiberg K. J. Chromatogr. A,2017,1481:116-126.
Han F C,Hu F,Gao Y D,Zhang C,Xi Y,Zhao Y J,Zhan M,Jin Z P,Zheng W W. J. Environ. Occup. Med. (韩凤婵,胡焚,高煜东,张超,奚晔,赵宜静,詹铭,金志萍,郑唯韡. 环境与职业医学),2019,36(10):9422-9428.
Determination of 26 Organophosphate Esters in Serum by Ultra High Performance Liquid Chromatography-Tandem Mass Spectrometry
In Situ Synthesis of Au/Foam Nickel Magnetic Substrate for Rapid Detection of Immunosuppressive Drug Metabolites in Serum by Using Surface Enhanced Raman Scattering Spectroscopy
Determination of Free Thyroid Hormones in Serum by Gel Filtration Chromatography Combined with Liquid Chromatography-Tandem Mass Spectrometry
Related Author
LIN Qiang
WANG Jian
LI Jing-jing
SU Dong-xia
LI Mei-li
WANG Jia
NIU Yu-min
NIU Yu-min
Related Institution
Beijing Yanqing District Center for Disease Prevention and Control
Yongning Community Health Service Center in Yanqing District
School of Public Health,Capital Medical University
Beijing Key Laboratory of Diagnostic and Traceability Technologies for Food Poisoning,Beijing Center for Disease Prevention and Control
Beijing Key Laboratory of Diagnostic and Traceability Technologies for Food Poisoning Beijing Center for Disease Prevention and Control