采用电泳沉积与自聚合方法在镍钛(NiTi)合金表面组装聚多巴胺(PDA)修饰的纳米/亚微米二氧化钛(PDA@TiO,2,NPs)固相微萃取(SPME)纤维涂层。电泳沉积前,先将NiTi合金水热处理原位生成氧化钛和氧化镍复合纳米片(TiO,2,NiOCNFs@NiTi),使涂层与NiTi合金牢固结合。将制备的PDA@TiO,2,NPs@TiO,2,NiOCNFs@NiTi纤维与高效液相色谱(HPLC)联用,用于富集、萃取和测定环境水样中的5种邻苯二甲酸酯(PAEs)类有机污染物,并考察了影响萃取效率的主要因素。在优化条件下,采用外标法对方法学参数进行评价。结果显示,5种PAEs在0.1 ~ 200 μg/L范围内呈良好的线性关系,相关系数(,r,)不小于0.998 9,检出限(LOD)和定量下限(LOQ)分别为0.013 ~ 0.055 μg/L和0.043 ~ 0.18 μg/L。单支纤维在日内和日间对50 μg/L PAEs标准溶液萃取测定的相对标准偏差(RSD)分别为5.4% ~ 6.8%和5.7% ~ 7.0%。该方法成功用于实际水样中痕量PAEs的测定,加标回收率为87.8% ~ 102%,RSD不大于8.2%。该方法灵敏度高、准确度好,适用于不同环境水样中痕量PAEs的高效分析测定。
A new polydopamine modified titanium dioxide nano/submicro particles coating(PDA@TiO,2,NPs) was fabricated on the NiTi alloy wire by electrophoretic deposition and self-polymerization modification in this paper.Prior to electrophoretic deposition,the titanium and nickel oxide composite nanoflakes(TiO,2,NiOCNFs) were in situ grown on NiTi alloy wire via hydrothermal treatment in order to boost the binding strength between TiO,2,NPs coating and NiTi fiber. Coupled with high performance liquid chromatography(HPLC),the fabricated PDA@TiO,2,NPs@TiO,2,NiOCNFs@NiTi fiber was applied as a solid phase microextraction(SPME) fiber to the determination of 5 phthalic acid esters(PAEs) in water samples.The main parameters affecting extraction efficiency such as extraction temperature,extraction time,stirring rate and ion strength were investigated and optimized.Under the optimized conditions,the analytical parameters including linear ranges,recoveries,limits of detection(LODs) and limits of quantitation(LOQs) were investigated and evaluated by external standard method.The results showed that the proposed method showed good linearity in the range of 0.1-200 μg/L,with correlation coefficients(,r,) not less than 0.998 9.LODs and LOQs were in the range of 0.013-0.055 μg/L and 0.043-0.18 μg/L,respectively.The intra-day and inter-day relative standard deviations(RSDs) for each fiber ranged from 5.4% to 6.8% and 5.7% to 7.0%,respectively.Moreover,the developed method was successfully used to extract and detect trace PAEs in real water samples.The spiked recoveries for target PAEs in real water samples ranged from 87.8% to 102%,with RSDs not more than 8.2%.In addition,the established method has high sensitivity,good accuracy,which was suitable for the high efficient detection of PAEs in different environmental water samples.
聚多巴胺电泳沉积纳米/亚微米二氧化钛固相微萃取高效液相色谱邻苯二甲酸酯
polydopamineelectrophoretic depositiontitanium dioxide nano/submicro particlessolid phase microextractionhigh performance liquid chromatographyphthalic acid esters
Gao D W,Li Z,Wen Z D,Ren N Q.Chemosphere,2014,95:24-32.
Li X X,Liu W J,Zhang C,Song P P,Wang J.Bull. Environ. Contam. Toxicol.,2020,104:301-306.
Le T M,Nguyen H M N,Nguyen V K,Nguyen A V,Vu N D,Yen N T H,Hoang A Q,Minh T B,Kannan K,Tran T M.Sci. Total Environ.,2021,788:147831.
Wang D X,Wang X C,Hu Q J,Wang F L,Li F.J. Instrum. Anal. (王东旭,王新财,胡奇杰,王凤丽,厉芬.分析测试学报),2022,41(10):1516-1522.
Wang X Y,Xue Y B,Zhe D M,Gao A,Guo R H,Gao J.China Plast. (王笑妍,薛燕波,者东梅,高昂,郭若 海,高静.中国塑料),2019,33(6):95-105.
Yao Y,Chen D Y,Yin J W,Zhou L,Cheng J Q,Lu S Y,Li H H,Wen Y,Wu Y.Environ. Int.,2020,143:105958.
Yang J L,Li Y X,Wang Y,Ruan J,Zhang J,Sun J C.TrAC Trends Anal. Chem.,2015,72:10-26.
GB 3838-2002. Environmental Quality Standards for Surface Water. National Standards of the People’s Republic of China(地表水环境质量标准.中华人民共和国国家标准).
Ortega-Zamora C,Jiménez-Skrzypek G,González-Sálamo J,Hernández-Borges J.J. Chromatogr. A,2021,1646:462132.
Ajdari B,Zahedi M M,Nassiri M,Jafari R.Water Sci. Technol.,2017,77:1782-1790.
Wu Y L,Zhou Q X,Yuan Y Y,Wang H Y,Tong Y Y,Zhan Y L,Sheng X Y,Sun Y,Zhou X Q.Talanta,2020,206:120213.
Li X H,Xiang L,Huang Y H,Mo C H,Li Y W,Li H,Cai Q Y.Chin. J. Anal. Chem. (李新洪,向垒,黄裕宏,莫测辉,李彦文,李慧,蔡全英.分析化学),2019,47(11):1842-1849.
Fernández-González V,Moscoso-Pérez C,Muniategui-Lorenzo S,LopezMahía P,Prada-Rodríguez D.Talanta,2017,162:648-653.
Feng Z M,Huang C H,Guo Y H,Tong P,Zhang L.Anal. Chim. Acta,2019,1084:43-52.
Wang X Q,Feng J J,Tian Y,Li C Y,Ji X P,Luo C N,Sun M.Talanta,2019,199:317-323.
Erdem P,Tağaç A A,Tashakkori P,Bozkurt S S,Merdivan M.Microchem. J.,2021,170:106648.
Li Y,Zhang M,Yang Y X,Wang X M,Du X Z.J. Chromatogr. A,2014,1358:60-67.
Wang F X,Zheng J,Qiu J L,Liu S Q,Chen G S,Tong Y X,Zhu F,Ouyang G F.ACS Appl. Mater. Interfaces,2017,9:1840-1846.
Rahmani F,Eshaghi A,Hosseini M M,Mollahosseini A.Microchem. J.,2019,145:942-950.
Cai Y,Yan Z H,Yang M,Huang X Y,Min W P,Wang L J,Cai Q Y.J. Chromatogr. A,2016,1478:2-9.
Huang Z Z,Lee H K.J. Chromatogr. A,2015,1414:41-50.
Klongklaew P,Bunkoed O.Microchem. J.,2021,165:106103.
Shakir S,Abdur -Rehman H M,Zahid R,Iwamoto M,Periasamy V.J. Alloys Compd.,2020,837:155579.
Xu H L,Li Y,Jiang D Q,Yan X P.Anal. Chem.,2009,81:4971-4977.
Lirio S,Fu C W,Lin J Y,Hsu M J,Huang H Y.Anal. Chim. Acta,2016,927:1-9.
0
浏览量
6
下载量
0
CSCD
关联资源
相关文章
相关作者
相关机构