建立了一种评价透析器对硫酸吲哚酚、硫酸对甲酸和马尿酸3种蛋白结合类尿毒症毒素(PBUTs)清除效果的高效液相色谱-三重四极杆质谱联用方法。配制与透析患者体内PBUTs浓度相当的模拟液代替血液,进行透析模拟治疗,取透析前后的模拟液作为待测样品。样品经乙腈沉淀蛋白和超滤离心处理后,采用Hypersil GOLD VANQUISH C,18,(100 mm × 2.1 mm,1.9 µm)色谱柱分离,以5 mmol/L乙酸铵溶液和甲醇为流动相进行梯度洗脱,负离子选择反应监测(SRM)模式扫描,内标法进行定量分析。3种PBUTs在25 ~ 1 000 ng/mL范围内线性关系良好,相关系数(,r,)均大于0.99,定量下限为25 ng/mL;在75、250、750 ng/mL加标水平下,回收率为92.5% ~ 114%,日内、日间相对标准偏差(RSD)均小于5.0%,基质效应为80.6% ~ 105%。该方法成本低、前处理简单、分析速度快、回收率高,适用于透析器对PBUTs清除效果的评价。
A method for evaluation on the removal effect of dialyzer to protein-bound uremic toxins(PBUTs) was established based on high performance liquid chromatography-triple quadrupole mass spectrometry.The dialysis simulated fluid with the same concentration of PBUTs as the dialysis patients’ blood was prepared instead of blood to conduct dialysis treatment in vitro due to the difficulty to obtain clinical blood samples and the complexity of patient blood composition,and the simulated fluid before and after dialysis was taken as the sample to be tested.After acetonitrile precipitation of protein and ultrafiltration centrifugation,the samples were separated on a Hypersil GOLD VANQUISH C,18,(100 mm × 2.1 mm,1.9 µm) column by gradient elution,using 5 mmol/L ammonium acetate solution and methanol as mobile phases.The mass spectrometric analysis was carried out under negative ion selective reaction monitoring mode(SRM),and the internal standard method was used for quantitative analysis of 3 PBUTs.The calibration curves of 3 PBUTs showed good linear relationships in the concentration range of 25-1 000 ng/mL,and the correlation coefficients(,r,) were more than 0.99.The limits of quantitation(LOQs) were 25 ng/mL.The average recoveries for 3 analytes at three added concentrations of 75,250 and 750 ng/mL ranged from 92.5%-114%,with the intra-day and inter-day RSDs both less than 5.0% and the matrix effects of 80.6%-105%.This method had the characteristics of low cost,simple pretreatment,short run-time and high accuracy,which provided a reliable technical support for evaluation on the removal effect of dialyzer to PBUTs.
蛋白结合类尿毒症毒素透析器模拟液高效液相色谱-三重四极杆质谱清除效果
protein-bound uremic toxinsdialyzersimulated fluidhigh performance liquid chromatography-triple quadrupole mass spectrometryremoval effect
International Society of Nephrology,International Federation of Kidney Foundations. Kidney Health for Everyone Everywhere-from Prevention to Detection and Equitable Access to Care. https://www.worldkidneyday.org/2020-campaign/2020-wkd-theme/https://www.worldkidneyday.org/2020-campaign/2020-wkd-theme/.
Chen X M,Ni Z H,Liu Y N,Xie Y S,Sun W.Chin. J. Intergr. Med. West. Med. (陈香美,倪兆慧,刘玉宁,谢院生,孙伟.中国中西医结合杂志), 2015,35(9):1029-1033.
Li L,Wang F,Xu M,Lu J L,Zhao Z Y,Li M,Wang T G,Wang S Y,Bi Y F,Xu Y,Cai W M,Ning G.Biomed. Environ. Sci.,2021,34(10):761-772.
Lu Y C,Wu C C,Tsai I T,Hung W C,Lee T L,Hsuan C F,Yu T H,Wei C T,Chung F M,Lee Y J,Wang C P.Clin. Chim. Acta,2021,516:83-91.
Tanaka H,Iwasaki Y,Yamato H,Mori Y,Komaba H,Watanabe H,Maruyama T,Fukagawa M.Bone,2013,56:347-354.
Tsai M H,Chang C H,Liou H H,Fang Y W.J. Clin. Med.,2021,10:744.
Watanabe H,Miyamoto Y,Otagiri M,Maruyama T.J. Pharm. Sci.,2011,100(9):3682-3695.
YY 0053-2016. Hemodialysis and Relevant Therapies-Haemodialysers,Haemodiafilters,Haemofilters and Haemoconcentrators. Pharmaceutical Industry Standard of the People’s Republic of China
血液透析及相关治疗 血液透析器、血液透析滤过器、血液滤过器和血液浓缩器.中华人民共和国医药行业标准.
Hideki K,Michio M,Hideki H,Tadao A.J. Jpn. Soc. Dial. Ther.,2012,45(5):435-445.
Zhao P,Chang J,Wu X J,Dong L P,Liu B J,Zhang Y F.J. Instrum. Anal. (赵鹏,常靖,吴小军,董林沛,刘冰洁,张云峰.分析测试学报),2022,41(2):227-233.
Niu Y M,Wang B,Yang R H,Duan H J,Zhang J,Shao B.J. Instrum. Anal. (牛宇敏,王彬,杨润晖,段鹤君,张晶,邵兵.分析测试学报),2020,39(6):715-721.
Zhang F,Yu Z R,Li M Z,Qiao B.J. Instrum. Anal. (张凡,于智睿,李明珠,乔斌.分析测试学报),2022,41(3):368-374.
Du Q Y,Zhang Y F,Wang J F,Wang L,Chang J,Dong L P,Wu X J,Liu B J.J. Instrum. Anal. 杜秋瑶,张云峰,王继芬,王璐,常靖,董林沛,吴小军,刘冰洁.分析测试学报),2020,39(4):485-491.
Fabresse N,Uteem I,Lamy E,Massy Z,Larabi I A,Alvarez J C.Clin. Chim. Acta,2020,507:228-235.
Itoh Y,Ezawa A,Kikuchi K,Tsuruta Y,Niwa T.Anal. Bioanal. Chem.,2012,403(7):1841-1850.
Lin C N,Wu I W,Huang Y F,Peng S Y,Huang Y C,Ning H C.J. Food Drug Anal.,2019,27(2):502-509.
Ma Y R,Xin M Y,Li K,Wang H,Rao Z,Liu T X,Wu X A.J. Pharm. Biomed. Anal.,2020,191:113551.
Prokopienko A J,West R E,Stubbs J R,Nolin T D.J. Pharm. Biomed. Anal.,2019,174:618-624.
Chang S,Chen X J,Xia T Y,Zhang F,Gao S H,Chen W S.Biomed. Chromatogr.,2016,30(11):1782-1788.
Wang Z P,Jiang H,Chen X J,Song X H,Xu F J,Chen F C,Mao Z G,Gao S H,Chen W S.J. Pharm. Biomed. Anal.,2020,186:113312.
Xu S W,Lin D Q,Yao S J.Acta Phys.-Chim. Sin. (徐诗文,林东强,姚善泾.物理化学学报),2016,32(7):1819-1828.
Deltombel O,Loor H,Glorieux G,Dhondt A,Biesen W V,Meijers B,Eloots S.Biochimie,2017,139:20-26.
Lu Z Y,Chen S X,Yao Y Y,Xing M M,Xie Y.Guangzhou Chem. Ind. (卢忠英,陈仕学,姚元勇,邢明明,谢勇.广州化工),2015,43(17):12-13.
Chinese Pharmacopoeia Commission. Chinese Pharmacopoeia. Beijing:China Medical Science and Technology Press(国家药典委员会.中华人民共和国药典.北京:中国医药科技出版社),2020.
0
浏览量
12
下载量
0
CSCD
关联资源
相关文章
相关作者
相关机构