王丰琳,周新颖,王文晶等.基于光源切换的发光二极管诱导荧光检测器的研制与评价[J].分析测试学报,2021,40(10):1460-1466.
WANG Feng-lin,ZHOU Xin-ying,WANG Wen-jing,et al.Development and Evaluation of a Light-emitting Diode Induced Fluorescence Detector Based on Light Source Switching[J].Journal of Instrumental Analysis,2021,40(10):1460-1466.
王丰琳,周新颖,王文晶等.基于光源切换的发光二极管诱导荧光检测器的研制与评价[J].分析测试学报,2021,40(10):1460-1466. DOI: 10.19969/j.fxcsxb.21031002.
WANG Feng-lin,ZHOU Xin-ying,WANG Wen-jing,et al.Development and Evaluation of a Light-emitting Diode Induced Fluorescence Detector Based on Light Source Switching[J].Journal of Instrumental Analysis,2021,40(10):1460-1466. DOI: 10.19969/j.fxcsxb.21031002.
荧光检测器具有灵敏度高、选择性好等特点,是高效液相色谱法(HPLC)常用的检测器之一。该文研制了一款基于共聚焦光学构型的多波长发光二极管(LED)诱导荧光检测器,可实现多种物质的检测。通过电机多重校准定位与光源滑动定位技术实现多光源的精准定位,使用共用发射光路简化仪器结构并降低了仪器成本。针对LED光源切换机构进行3万个周期连续切换的可靠性测试,测得激光器的靶点位置误差在 ± 0.2 mm以内,实现了LED光源定位的准确性和良好重复性,解决了传统荧光检测器结构复杂、价格高,以及发光二极管(LED)单波长诱导荧光检测器专用性强、应用范围窄等问题。以甲醇为流动相,使用C,18,色谱柱对3个光源进行基线考察,结果表明:研制的检测器在340、365、385 nm激发光源下,基线噪声低于5.0 × 10,-4, FU,漂移小于5.0 × 10,-3, FU/h。以真菌毒素和苯并(α)芘为探针,对其线性和灵敏度进行评价。分别采用乙腈-水-乙酸(96∶102∶2,体积比)、甲醇-乙腈-水(20∶20∶60)、乙腈-水(90∶10)作为检测赭曲霉素A、黄曲霉毒素B,1,、苯并(α)芘的流动相,激发波长分别为340、365、385 nm。测得赭曲霉素A、黄曲霉毒素B,1,、苯并(α)芘分别在1.0~50.0、0.1~20.0、0.5~20.0 ng/mL质量浓度范围内呈良好的线性关系(,r,2, ≥ 0.999 9),检出限分别为0.06、0.01、0.02 ng/mL。对该检测器的重复性进行考察,结果表明3个光源对样品响应值的RSD(,n, = 11)均小于1.0%。以340 nm光源为测试条件,显示检测器在震动、高温和低温环境下的基本性能均未发生明显变化。与市售进口的通用型荧光检测器相比,二者在基线噪声、漂移和灵敏度方面结果相当。该文研制的检测器结构简单、波长覆盖范围宽、性能良好,在真菌毒素和苯并(α)芘检测中具有良好的应用前景。
Fluorescence detector has the characteristics of high sensitivity and good selectivity.It is one of the most commonly used detectors in high performance liquid chromatography(HPLC).In this paper,a multi-wavelength light-emitting diode(LED) induced fluorescence detector was developed based on a confocal optical configuration,which could be applied in detection of variety of substances.The motor multiple calibration positioning technique and light source sliding positioning technique were used to achieve the precise positioning of multiple light sources.And a shared emission optical path was used to simplify the instrument structure and reduce the cost of the instrument.The reliability test on the continuous switching of LED light source for 30 000 cycles showed that the target position errors of the laser ranged within ± 0.2 mm,which realized the accuracy and good repeatability of LED light source positioning,solving the problems of traditional fluorescence detector with complex structure and high price,and LED single wavelength induced fluorescence detector with strong specificity and narrow application range.A mobile phase of methanol and a C,18, chromatographic column were used to investigate the baselines of three light sources.Results showed that the developed detector,under excitation light sources of 340,365 and 385 nm ,had baseline noises less than 5.0 × 10,-4, FU and a drifts less than 5.0 × 10,-3, FU/h.Furthermore,mycotoxins and benzo(α)pyrene were used as probes to evaluate the linearity and sensitivity.Meanwhile,acetonitrile-water-acetic acid(96∶102∶2,by volume),methanol-acetonitrile-water (20∶20∶60) and acetonitrile-water(90∶10) were used as mobile phases for the analysis of ochratoxin A,aflatoxin B,1, and benzo(α)pyrene,respectively.The excitation wavelengths of the detector were set at 340,365 and 385 nm,respectively.The concentrations of ochratoxin A,aflatoxin B,1, and benzo(α)pyrene showed good linearities (,r,2 , ≥ 0.999 9) in the ranges of 1.0-50.0 ng/mL,0.1-20.0 ng/mL and 0.5-20.0 ng/mL,respectively.The limits of detection were 0.06,0.01 and 0.02 ng/mL,respectively.The repeatability of this detector was investigated,and the result showed that the RSDs (,n,=11) for sample response values of the three light sources were all less than 1.0%.While using the 340 nm light source as the test condition,the basic performance of the detector under vibration,high temperature and low temperature environments did not changed significantly.The tested results of the detector were comparable in terms of baseline noise,drift and sensitivity,compared with those of the commercially available imported general-purpose fluorescence detectors.With the advantages of simple structure,wide wavelength coverage and good performance,the developed detector showed good application prospects in the detection of mycotoxins and benzo(α)pyrene.
发光二极管(LED)多波长荧光检测器赭曲霉毒素A黄曲霉毒素B1苯并(α)芘
light-emitting diode (LED)multi-wavelengthfluorescence detectorochratoxin Aaflatoxin B1benzo(α)pyrene
Zhou W P,Cheng M L,Xu Z,Zhou S H,Ding L.J. Instrum. Anal. (周雯鹏,陈茂龙,许宙,朱绍华,丁利.分析测试学报),2020,39(2):1137-1142.
Yu X Q,Li S C,Tang C Y.J. Instrum. Anal. (余晓琴,李澍才,唐昌云.分析测试学报),2019,38(10):1260-1264.
Chen K,Zhang P,Wang Y,Zheng Z J,Dong J L,Sun L P,Zhu S Y,Zhao X E.J. Instrum. Anal. 陈坤,张鹏,王妍,郑振佳,董京磊,孙鲁平,朱树芸,赵先恩.分析测试学报),2020,39(3):301-307.
Chen C M,Lee J A,Huang T C.J. Biomed. Chromatogr.,2012,26(2):256-260.
Xue M,Zhao L H,Sun Y S,Tang T,Xia M Z.J. Anal. Sci. (薛敏,赵连海,孙元社,唐涛,夏明珠.分析科学学报),2014,30(2):269-272.
He C H,Wu J,Wang T H,Yang H Q.J. Anal. Sci. (何崇慧,吴晶,王廷海,杨红强.分析科学学报),2015,31(4):554-559.
Geng X H,Shi M,Ning H I,Feng C B,Guan Y F.Talanta,2018,76(1):279-284.
Jozef S,Josef P,Vladislav K.Anal. Chim. Acta,2020,1112:80-91.
Xu J,Xiong Y,Chen S H. Talanta,2008,75(4):885-889.
Li H F,Chen Z F,Lin J M.Chin. J. Anal. Chem. (李海芳,陈志峰,林金明.分析化学),2008,36(11):1587-1591.
Liu F,Yu S X,Tang T,Sun Y S,Zhang W B,Li T.Chin. J. Chromatogr. (刘翻,于叔新,唐涛,孙元社,张维冰,李彤.色谱),2011,29(9):896-900.
Liu J,Liu C C.Instrum. Technol. Sens. 刘剑,刘常春.仪表技术与传感器),2010,(8):17-18.
Geng X H,Gao Y,Feng C B,Guan Y F.Talanta,2017,22(7):183-188.
Yang B C,Gang Y F,Tan F.Prog. Chem. (杨丙成,关亚风,谭峰.化学进展),2004,16(6):871-878.
Zhong K K.Study on Fluorescence Detection System of Fiber Optic Quantitative PCR Instrument.Hangzhou:Zhejing University(钟珂珂.光纤型定量PCR仪荧光检测系统研究.杭州:浙江大学),2013.
Xi Q Y.Study on Ultra-sensitive Laser Induced Fluorescence Detector.Dalian: Dalian University of Technology(席秋颖.超高灵敏度激光诱导荧光检测器的研究.大连:大连理工大学),2020.
JJG 705-2014.National Metrology Regulations of the People's Republic of China - Liquid Chromatography.National Standard of the People's Republic of China(中华人民共和国国家计量规程 液相色谱仪.中华人民共和国国家标准).
GB/T 26792-2011.High Performance Liquid Chromatography.National Standard of the Peopie's Republic of China(高效液相色谱仪.中华人民共和国国家标准).
GB 5009.96-2016.Determination of Ochratoxin A in Food.National Standards of the People's Republic of China(食品中赭曲霉毒素A的测定.中华人民共和国国家标准).
GB 5009.22-2016.Determination of Aflatoxin B and G in Food.National Standards of the People's Republic of China(食品中黄曲霉毒素B族和G族的测定.中华人民共和国国家标准).
GB 5009.27-2016.Determination of Benzo(a)pyrene in Food.National Standards of the People's Republic of China(食品中苯并(a)芘的测定.中华人民共和国国家标准).
0
浏览量
4
下载量
0
CSCD
关联资源
相关文章
相关作者
相关机构