Determination of Glyphosate by High Performance Liquid Chromatography withο-Nitrobenzenesulfonyl Chloride as Derivatization Reagent[J]. 2011,30(6):683-686.
Determination of Glyphosate by High Performance Liquid Chromatography withο-Nitrobenzenesulfonyl Chloride as Derivatization Reagent[J]. 2011,30(6):683-686.DOI:
A derivatization method for the determination of glyphosate was developed by using o nitrobenzenesulfonyl chloride as derivatization reagent.The optimized conditions of derivatization were as follows:reaction temperature:25 ℃,reaction time:10 min,buffer solution:0.25 mol/L borate buffer(pH 9.0),the molar ratio of glyphosate to o-nitrobenzenesulfonyl:1∶5.The optimal chromatographic conditions were as follows:Lichrospher C18 column(5 μm,4.6 mm×250 mm),detection wavelength:230 nm,flow rate:1.0 mL/min,mobile phase:methanol-pH 5.5 0.05 mol·L-1phosphate buffer(20∶80).Under the optimal conditions,the calibration curve for glyphosate was linear in the range of 0.05-1.0 g·L-1 with a correlation coefficient (r2)of 0.999 3.The limit of detection was 6.25 mg·L-1.The method was simple,rapid and reliable,and was suitable for the determination of glyphosate in general laboratory.