Simultaneous Determination of 14 Macrolide Drugs Residues in Pork Samples by Ultra Performance Liquid Chromatography-Tandem Mass Spectrometry with Solid Phase Extraction
|更新时间:
|
Simultaneous Determination of 14 Macrolide Drugs Residues in Pork Samples by Ultra Performance Liquid Chromatography-Tandem Mass Spectrometry with Solid Phase Extraction
王一涵, 王展华, 陈万勤, 王峰, 刘柱. Simultaneous Determination of 14 Macrolide Drugs Residues in Pork Samples by Ultra Performance Liquid Chromatography-Tandem Mass Spectrometry with Solid Phase Extraction[J]. 2019, 38(10): 1247-1253.
DOI:
王一涵, 王展华, 陈万勤, 王峰, 刘柱. Simultaneous Determination of 14 Macrolide Drugs Residues in Pork Samples by Ultra Performance Liquid Chromatography-Tandem Mass Spectrometry with Solid Phase Extraction[J]. 2019, 38(10): 1247-1253.DOI: doi:10.3969/j.issn.1004-4957.2019.10.014.
Simultaneous Determination of 14 Macrolide Drugs Residues in Pork Samples by Ultra Performance Liquid Chromatography-Tandem Mass Spectrometry with Solid Phase Extraction
An ultra performance liquid chromatography-tandem mass spectrometric method with solid phase extraction was established for the simultaneous determination of 14 macrolide drugs in pork samples.Samples were extracted with 70%(volume fraction) acetonitrile
purified by solid phase extraction using a hydrophilic lipophilic balanced(HLB) polymer
and separated on a C18 column(100 mm×2.1 mm,1.7 μm) with acetonitrile and 01% formic acid-2 mmol/L ammonium acetate solution as mobile phases by gradient elution.The analytes were detected by mass spectrometry in positive ion mode under multiple reaction monitoring(MRM) scanning mode
and quantified by external standard method.Results showed that good linearities for 14 macrolide drugs were obtained in the corresponding mass concentration ranges with correlation coefficients(r2) larger than 0.995.The detection limits and quantitation limits of the method were in the ranges of 0.1-0.8 μg/kg and 0.5-3.0 μg/kg
respectively.Recoveries at low
medium and high spiked levels were in the range of 71.1%-102%
with relative standard deviations(RSD) less than 6.0%.This method could be applied in the analysis of macrolide drugs in pork samples with good accuracy and reliability.
Detection of Sodium Trithiocyanurate in Wheat Flour and Flour Additives by Ultra High Performance Liquid Chromatography-Tandem Mass Spectrometry with Pass through SPE Cleanup
Residue Analysis of Mefentrifluconazole and Its Four Metabolites in Dendrobium officinale by Ultra Performance Liquid Chromatography-Tandem Mass Spectrometry
Determination of 9 Alkylphenols and Alkylphenol Ethoxylates in Down by High Performance Liquid Chromatography-Tandem Mass Spectrometry Combined with Solid Phase Extraction
Definition Clarification and Detection Reliability Analysis of Environmental Micro/Nanoplastics
Advances in the Formation Mechanisms,Characterization Methods,and Toxicological Effects of Surface Coronas on Micro/Nanoplastics