Rapid Determination of 17 Quinolones in Cosmetics Using Ultra performance Liquid Chromatography-Quadrupole/Electrostatic Field Orbitrap High Resolution Mass Spectrometry
廖华勇, 王景, 李红玉. Rapid Determination of 17 Quinolones in Cosmetics Using Ultra performance Liquid Chromatography-Quadrupole/Electrostatic Field Orbitrap High Resolution Mass Spectrometry[J]. 2019, 38(9): 1102-1107.
廖华勇, 王景, 李红玉. Rapid Determination of 17 Quinolones in Cosmetics Using Ultra performance Liquid Chromatography-Quadrupole/Electrostatic Field Orbitrap High Resolution Mass Spectrometry[J]. 2019, 38(9): 1102-1107.DOI: doi:10.3969/j.issn.1004-4957.2019.09.012.
An ultra performance liquid chromatography-quadrupole/electrostatic field orbitrap high resolution mass spectrometry(UPLC-Q Orbitrap HRMS) was established for the determination of 17 quinolones in lotions
emulsions and non wax based cream cosmetics.The samples were ultrasonically extracted with 0.1%formic acid and acetonitrile mixed solvent
purified with n hexane
and then separated with an Agilent Poroshell EC C18(4.6 mm×150 mm,2.7 μm) column.The analytes were detected in Full Scan/dd-MS2 mode
and quantified by the internal standard method.Seventeen quinolones showed good linearities in the range of 0.05-200 μg/L with their correlation coefficients(r2) larger than 0.996 0.The limits of detection and quantitation for 17 quinolones were 0.5 μg/kg and 1.0 μg/kg
respectively.Average recoveries for 17 quinolones at three spiked levels of 2,4,20 μg/kg ranged from 82.3%to 108%with relative standard deviations(RSD,n=6) of 3.5%-8.6%.The method is rapid
accurate and sensitive
and is suitable for the determination of quinolone antibiotics in cosmetics.