Determination of Organophosphate Esters in Medical Masks by Ultra-high Performance Liquid Chromatography-Tandem Mass Spectrometry
In this study,a method was developed for simultaneous determination of 16 organophosphate esters(OPEs) in medical masks by ultra-high performance liquid chromatography-tandem mass spectrometry(UPLC-MS/MS).In addition,the extraction and purification process for 16 OPEs was optimized.Specifically,three different extracting solvents including dichloromethane-,n,-hexane(1∶1),dichloromethane-acetonitrile(1∶1) and methanol,coupled with two purification processes,were compared. Meanwhile,the mask samples were ultrasonically extracted three times with methanol,then purified with ENVI-18 cartridges using 6 mL acetonitrile containing 25% dichloromethane as the eluent.The analytes were separated on a Acquity UPLC BEH C,18, column by gradient elution using 5 mmol/L ammonium acetate and methanol as the mobile phases,then analyzed in electrospray ionization positive(ESI+) mode under the multiple reaction monitoring(MRM) mode,and finally quantified by internal standard method.Results indicated that the limits of detection(LOD,,S/N, = 3) for 16 OPEs in mask were in the range of 0.004 60-1.03 ng/dm,2,,while the limits of quantitation(LOQ,,S/N, = 10) were 0.015 2-3.43 ng/dm,2,.The average recoveries for 16 OPEs in medical mask at three spiked levels of 5, 20 and 40 ng ranged from 68.8% to 140%,with relative standard deviations(RSD) of 0.70%-18%.These results indicated that the established method was suitable for the determination of OPEs in medical mask samples.The method was applied to the quantitation of 16 OPEs in forty-two medical mask samples.The total concentrations of 16 OPEs were in the range of 6.60-2 387 ng/dm,2,,with a median concentration of 34.2 ng/dm,2,.The detection frequencies for twelve OPEs were higher than 50%.Tri-phenyl phosphate(TPHP) and tri(1-chloro-2-propyl) phosphate(TCIPP) were the most abundant OPEs,with concentrations of 0.131-2 274 ng/dm,2, and 0.370-79.8 ng/dm,2,,respectively.Considering that the concentration of each analyte was relatively low,it is speculated that the OPEs in medical mask may be adsorbed from the air or transferred from plastic packaging during the manufacturing or packaging process.
关键词
有机磷酸酯超高效液相色谱-串联质谱医用口罩纺织品阻燃剂
Keywords
organophosphate esterultra-high performance liquid chromatography-tandem mass spectrometrymedical masktextileflame retardant
references
Wei G L,Li D Q,Zhuo M N,Liao Y S,Xie Z Y,Guo T L,Li J J,Zhang S Y,Liang Z Q.Environ. Pollut.,2015,196:29.
Stapleton H M,Sharma S,Getzinger G,Ferguson P L,Gabriel M,Webster T F,Blum A.Environ. Sci. Technol.,2012,46(24):13432.
van der Veen I,de Boer J.Chemosphere,2012,88(10):1119-1153.
Li W H,Wang Y,Asimakopoulos A G,Covaci A,Gevao B,Johnson-Restrepo B,Kumosani T A,Malarvannan G, Moon H B,Nakata H,Sinha R K,Tran T M,Kannan K.Environ. Int.,2019,133:105178.
Cequier E,Ionas A C,Covaci A,Marce R M,Becher G,Thomsen C.Environ. Sci. Technol.,2014,48(12):6827-6835.
Li J H,Zhao L M,Letcher R J,Zhang Y Y,Jian K,Zhang J H, Su G Y.Environ. Int.,2019,127:35-51.
Zhang Q,Lu M Y,Dong X W,Wang C,Zhang C L,Liu W P,Zhao M R.Environ. Sci. Technol.,2014,48(12):6995-7001.
Zhang Q,Ji C Y,Yin X H,Yan L,Lu M Y,Zhao M R.Environ. Pollut.,2016,210:27-33.
Luo D,Liu W Y,Tao Y,Wang L M,Yu M,Hu L Q,Zhou A F,Covaci A,Xia W,Li Y Y,Xu S Q,Mei S R.Environ. Sci. Technol.,2020,54(6):3375-3385.
Dishaw L V,Powers C M,Ryde I T,Roberts S C,Seidler F J,Slotkin T A,Stapleton H M.Toxicol. Appl. Pharmacol.,2011,256(3):281-289.
Wang J,Pan L J,Tang S,Ji J S,Shi X M.Environ. Pollut.,2020,266:115099.
Li G Z,Li G M,Li Y H,Huang J Y,Zhao J Y.Text. Sci. Res. (李冠志,李桂梅,李昱昊,黄景莹,赵瑾瑜.纺织科学研究),2021,2:41-43.
Zhang N,Zhang B,Yu W T,Li X,Mai B H,Yu W J.Chin. J. Anal. Lab. (张楠,张斌,禹伟腾,李宣,麦宝华,于文杰.分析试验室),2015,34(12):1411-1414.
Ma L,Chen M,Bai H,Bai D,Li M,Ma Q.J. Chin. Mass Spectrom. Soc. (马麟,陈萌,白桦,柏冬,李明,马强.质谱学报),2020,41(6):556-568.
Zhu H,Al-Bazi M M,Kumosani T A,Kannan K.Environ. Sci. Technol. Lett.,2020,7(6):415-420.
GB 19083-2010.Technical Requirements for Protective Face Mask for Medical Use.National Standards of the People's Republic of China(医用防护口罩技术要求.中华人民共和国国家标准).
YY 0469-2011.Surgical Mask.Pharmaceutical Industry Standards of the People's Republic of China(医用外科口罩.中华人民共和国医药行业标准).
Hou M M,Shi Y L,Cai Y Q.Chin. J. Chromatogr. (侯敏敏,史亚利,蔡亚岐.色谱),2021,39(1):69-76.
Determination of Organophosphate Esters and Their Diester Metabolites in Cereals by Ultra⁃high Performance Liquid Chromatography-Tandem Mass Spectrometry
Simultaneous Determination of 15 Coumarins and 10 Benzophenones in Sunscreen Cosmetics by QuEChERS/Ultra-high Performance Liquid Chromatography-Tandem Mass Spectrometry
Determination of Tacrolimus in Plasma by Supramolecular Solvent Extraction/Ultra-high Performance Liquid Chromatography- Tandem Mass Spectrometry
Determination of 24 Kinds of Phenolic Acids and Flavonoids in Honey by Ultra-high Performance Liquid Chromatography-Tandem Mass Spectrometry
Simultaneous Determination of the Migration of 20 Photo Initiators in Food Contact Inks by UPLC-MS/MS
Related Author
No data
Related Institution
School of Environment,Hangzhou Institute for Advanced Study,UCAS
State Key Laboratory of Environmental Chemistry and Ecotoxicology,Research Center for Eco⁃Environmental Science,Chinese Academy of Sciences
Beijing Key Laboratory of Nutrition & Health and Food Safety, Nutrition & Health Research Institute,COFCO Corporation
National Liquor Product Quality Supervision and Inspection Center