MA Meng-yao,AN Shao-hang,LIU Shu-jiang,SHENG Yu-qiang,CHANG Yin-zhong,CHEN Zhan-ying,LI Qi,WANG Shi-lian.Research on the Method for Oxygen and Argon Separation and Argon Quantitative Analysis Based on Gas Chromatography[J].Journal of Instrumental Analysis,2025,44(11):1-5. DOI: 10.12452/j.fxcsxb.25021489.
Research on the Method for Oxygen and Argon Separation and Argon Quantitative Analysis Based on Gas Chromatography
Ar是全面禁止核试验条约现场视察监测的重要核素,在氩总量定量分析时,样品气中残留的微量氧气会干扰氩的测量结果。针对氧氩分离分析难题,基于气相色谱技术,实验研究了色谱柱箱温度、进样口压力、氧氩混合气浓度对氧氩分离效果的影响。结果表明,采用HP-5A分子筛色谱柱,在柱箱温度为-65 ℃,进样口压力为35 kPa时,可实现不同浓度氧氩的分离分析。在该条件下,50% O
clear explosions,is a key Comprehensive Nuclear-Test-Ban Treaty(CTBT) verification indicator. Its activity concentration is measured against collected argon volume; levels above background indicate a nuclear test. However,residual oxygen in samples,which has nearly identical physical properties to argon,is difficult to remove completely,causing errors in argon volume and
37
Ar quantification,compromising verification reliability and potentially leading to false compliance conclusions
.
To address this oxygen-argon separation analytical challenge,this study systematically investigated the separation efficiency of oxygen and argon mixtures using gas chromatography(GC). Key operational parameters including column oven temperature,inlet pressure,and gas concentration were optimized to achieve the baseline separation of oxygen and argon. After an extensive series of experiments and in-depth analysis,the results demonstrated that a column oven temperature of -65 ℃ and an inlet pressure of 35 kPa could effectively separate oxygen and argon across a wide range of concentration ratios. For a 50% O
2
-50% Ar mixture,a resolution(
R
s
) of 1.50 was achieved,indicating a good separation degree. While for a 95% O
2
-5% Ar mixture,although the separation was more challenging,an
R
s
value of 1.14 was still obtained,and in both cases,the total analysis time were within 5 minutes,which demonstrate the high efficiency of this method. Furthermore,a linear external standard calibration curve with a high correlation coefficient(
r
2
>
0.999) was established for argon quantification,validated using an oxygen-rgon mixture gas with an argon volume ratio concentration of 3%. Verification tests were conducted,and the results showed that the relative deviation between the experimentally measured argon concentrations and the certified reference values was less than 4%. This low relative deviation not only highlights the hig
h precision of the established method but also provides strong support for its reliability in practical applications,ensuring more accurate and reliable nuclear test verification under the CTBT framework.
关键词
Keywords
references
Riedmann R A , Purtschert R . Environ. Sci. Technol. , 2011 , 45 ( 20 ): 8656 - 8664 .
Saldanha R , Back H O , Tsang R H M , Alexander T , Elliott S R , Ferrara S , Mace E , Overman C , Zalavadia M . Phys. Rev. C , 2019 , 100 ( 2 ): 024608 .
Purtschert R , Riedmann R , Loosli H H . 4th Mini Conference on Noble Gases in the Hydrosphere and in Natural Gas Reservoirs , 2007 : 38 .
Aalseth C E , Day A R , Haas D A , Hoppe E W , Hyronimus B J , Keillor M E , Mace E K , Orrell J L , Seifert A , Woods V T . Nucl. Instrum. Methods Phys. Res.,Sect. A , 2011 , 652 ( 1 ): 58 - 61
Li W , Gong J , Bian Z S . NuclChina. Sci. Technol. Rep. (李伟 , 龚建 , 卞直上 . 中国核科技报告) , 2005 ,( 2 ): 90 - 101 .
Duan R L , Liu S P , Yang X B , Chen Y L , Gong J , Liu S X , Yao Q Y , Xiong Z H . At. Energy Sci. Technol. (段荣良,刘素萍,杨晓波,陈银亮,龚健,刘书行,姚琼英,熊宗华. 原子能科学技术), 2000 , 34 ( 3 ): 270 - 270 .
Li Q , Fan Y Q , Wang S L , Zhao Y G , Chang Y Z , Jia H M , Liu S J , Zhang X J . At. Energy Sci. Technol. (李奇,樊元庆,王世联,赵允刚,常印忠,贾怀茂,刘蜀疆,张新军. 原子能科学技术), 2019 , 53 ( 4 ): 734 - 738 .
Haas D A . The Science Case for 37 Ar as a Monitor for Underground Nuclear Explosions,2010 .[ 2025-03-28 ]. https://www.pnnl.gov/main/publications/external/technical_reports/PNNL-19458.pdf https://www.pnnl.gov/main/publications/external/technical_reports/PNNL-19458.pdf .
Li J R , Kuppler R J , Zhou H C . Chem. Soc. Rev. , 2009 , 38 : 1477 - 1504 .
Ma M Y , Liu S J , Sheng Y Q , An S H , Chang Y Z , Chen Z Y , Li Q , Wang S L . J. Nucl. Radiochem. (马梦瑶,刘蜀疆,盛毓强,安少杭,常印忠,陈占营,李奇,王世联. 核化学与放射化学), 2024 , 46 ( 5 ): 425 - 436
Zhou G D . Dictionary of Chemistry . 2nd ed . Beijing : Chemical Industry Press(周公度. 化学辞典. 2版. 北京:化学工业出版社) , 2004 : 771 - 772 .
Karlsson B M . Anal. Chem. , 1966 , 38 ( 4 ): 440 - 443 .
DALIAN Special Gases . China Patent (大连大特气体有限公司 . 中国专利) ,201320887606.2. [ 2014-07-02 ].
Fu Z J . China Patent (付振江 . 中国专利) ,201721163639.7. [ 2018-05-08 ].
Lard E W , Horn R C . Anal. Chem. , 1960 , 32 ( 7 ): 878 - 879 .
Chen Z Y , Zhang H T , Wang J , Chang Y Z . Chin. J. Anal. Lab. (陈占营 , 张海涛 , 王军 , 常印忠 . 分析试验室) , 2008 ,( S1 ): 315 - 317 .
Chen Z Y , Liu S J , Wang J L , Chang Y Z . Chin. J. Anal. Chem. (陈占营,刘蜀疆,王建龙,常印忠. 分析化学), 2016 , 44 ( 3 ): 468 - 473 .
Chen Z Y , Liu S J , Chang Y Z , Wang J L , Hei D W . Mod. Appl. Phys. (陈占营,刘蜀疆,常印忠,王建龙,黑东炜. 现代应用物理), 2024 , 15 ( 5 ): 155 - 163 .
Chen Z Y , Liu S J , Wang J L , Chang Y Z , Fan Y Q , Li Q . At. Energy Sci. Technol. (陈占营,刘蜀疆,王建龙,常印忠,樊元庆,李奇.原子能科学技术), 2016 , 50 ( 11 ): 1949 - 1955 .
Liu G Q , Yu Z L . Column Technique . 2nd ed . Beijing : Chemical Industry Press(刘国诠,余兆楼. 色谱柱技术. 2版. 北京:化学工业出版社) , 2006 : 94 - 95 .
Liu S J , Chen Z Y , Chang Y Z , Wang S L , Li Q , Shi J F , Fan Y Q , Zhao Y G . At. Energy Sci. Technol. (刘蜀疆,陈占营,常印忠,王世联,李奇,石建芳,樊元庆,赵允刚. 原子能科学技术), 2012 , 46 : 115 - 119 .
Han J T . J. Xinyang Norm . Univ. : Nat. Sci. Ed . (韩金土. 信阳师范学院学报:自然科学版), 1990 , 3 ( 4 ): 351 - 354 .
Chen X F . Chemistry (陈祥舫 . 化学通报) , 1982 ,( 12 ): 61 - 62 .
Shi B Q , You T . InstrumAnal. (石宝清 , 由田 . 分析仪器) , 2003 ,( 3 ): 38 - 39 .
Rege S U , Yang R T . Adsorption , 2000 , 6 ( 1 ): 15 - 22 .
Wang Y . Sep. Purif. Technol. , 2023 , 312 : 123372 .
Xiong Y J , Wang C Y , Wei R Y , Sun X Q . Acta Ecol. Anim. Domast. (熊一键,王晨昱,魏冉月,孙小琴. 家畜生态学报), 2020 , 41 ( 2 ): 22 - 28 .
Du Y Y , Meng X J , Yang B , Song L , Zhu G X , Zhou X , Zhang A P , Pan J , Jiang L L . Chin. J. Mass Spectrom. Soc. , 2022 , 43 ( 4 ): 512 - 521 .
Simultaneous Determination of 11 Odorants in Drinking Water by Liquid-Liquid Extraction/Gas Chromatography-Quadrupole Electrostatic Field Orbitrap Mass Spectrometry
Application of T-C Composite Adsorption Tube in the Determination of Low-boiling Volatile Organic Compounds in Indoor Air
Development and Application of Online Monitoring System for Odor Pollutants
Related Author
DU Ming-yue
ZHAN Wei
CHEN Zhe
ZHANG Yong-xin
ZHANG Jing
LIANG Ji-pan
YU Tang
JIANG Wei-wu
Related Institution
Key Laboratory of Urban Stormwater System and Water Environment,Ministry of Education,Beijing University of Civil Engineering and Architecture
School of Public Health,China Medical University
Beijing Center for Disease Prevention and Control
The Construction Quality Supervision and Inspection Station of Zhuhai
Guangdong Academy of Building Research Group Co.,Ltd