Rapid Determination of Memantine Hydrochloride in Memantine Hydrochloride Tablets by Contactless Conductivity Method with Microfluidic Chip
|更新时间:2023-05-25
|
Rapid Determination of Memantine Hydrochloride in Memantine Hydrochloride Tablets by Contactless Conductivity Method with Microfluidic Chip
Vol. 39, Issue 2, Pages: 263-267(2020)
作者机构:
1. 南方医科大学珠江医院药学部
2. 深圳市第二人民医院药学部
3. 中山大学药学院
作者简介:
基金信息:
DOI:
CLC:
扫 描 看 全 文
Rapid Determination of Memantine Hydrochloride in Memantine Hydrochloride Tablets by Contactless Conductivity Method with Microfluidic Chip. [J]. 39(2):263-267(2020)
DOI:
Rapid Determination of Memantine Hydrochloride in Memantine Hydrochloride Tablets by Contactless Conductivity Method with Microfluidic Chip. [J]. 39(2):263-267(2020)DOI:
Rapid Determination of Memantine Hydrochloride in Memantine Hydrochloride Tablets by Contactless Conductivity Method with Microfluidic Chip
A contactless conductivity method with microfluidic chip was established for the rapid determination of memantine hydrochloride in memantine hydrochloride tablet.Effects of electrophoresis parameters such as variety,concentration,pH value of buffer solution,injection time and separation voltage on the separation detection were investigated.The optimized buffer system was 3.0 mmol/L triethylamine-2.0 mmol/L phosphate buffer solution(pH 3.3) containing 2%(by volume) dimethyl sulfoxide(DMSO),the injection time was 10 s and the separation voltage was 2.0 kV.In the optimal conditions,the linear range for memantine hydrochloride was 10-2 000 μg/mL(r2=0.999 1).The limits of detection and quantitation were 7 μg/mL and 10 μg/mL,respectively.Relative standard deviations for precision,stability and reproducibility tests were lower than 2.0%,and average recoveries for memantine hydrochloride were in the range of 96.5%-99.2%,with a detection time less than 18 s.The method is rapid and simple,and is suitable for the determination of memantine hydrochloride in memantine hydrochloride tablet.