Preparation and Characterization of a Mercury(Ⅱ) selective Polymeric Membrane Electrode. [J]. 30(2):207-212(2011)
高产率合成了一种新的Schiff base结构化合物,并将其表征为高选择性聚合物膜汞离子选择性电极载体。考察了不同增塑剂及离子交换剂对膜电极响应性能的影响,在最佳膜组分条件下测得该电极对汞离子的线性响应范围为10×10-6~30×10-4 mol/L,响应斜率为(293±03) mV/dec,检出限为26×10-7 mol/L;该电极响应速率快(小于12 s),可在较宽的pH范围内(pH 28~56)使用,且其它常见碱金属、碱土金属以及过渡金属离子对该测试电极的干扰小;可准确检测自来水中汞离子的浓度。
A new Schiff base compound derived from 1,2,4 triazole was synthesized with a high yield.Potentiometric responses of poly(vinyl) chloride(PVC) electrode based on this ligand as neutral ionophore showed high sensitivity and selectivity towards Hg2+ ion.Effects of different plasticizers and amounts of ion exchanger on the potentiometric response property of Hg2+ selective electrode(Hg2+-ISE) were discussed.The result indicated that,under the optimal membrane composition [PVC:3270,o NPOE:6539,ionphore:099,ion exchanger:092,(w∶w)],a good Nernstian response slope of (293±03) mV/dec was obtained for Hg2+ in a wide concentration range of 10×10-6-30×10-4 mol/L with a detection limit of 26×10-7 mol/L Except for silver ion,most of alkali metal,alkaline earth metal and other common transition metal ions didnt interfere the performance of the membrane electrode.The fabricated Hg2+-ISE can be used in a wide pH range of 28-56 with a fast response time of less than 12 s and can be used for 3 month without any obvious change in potentiometric response.The proposed electrode was also successfully applied as indicator electrode in the potentiometric detection of Hg2+ in lab water by standard addition method and the results were confirmed by ICP-MS.
离子选择性电极汞离子Schiff base
ion selective electrodemercury ion(Ⅱ)Schiff base
0
Views
892
下载量
2
CSCD
Publicity Resources
Related Articles
Related Author
Related Institution
京公网安备11010802024621