Determination of Benzo(a)pyrene and Bisphenol A in Drinking Water by Solid Phase Extraction/High Performance Liquid Chromatography
|更新时间:
|
Determination of Benzo(a)pyrene and Bisphenol A in Drinking Water by Solid Phase Extraction/High Performance Liquid Chromatography
Vol. 34, Issue 7, Pages: 848-851(2015)
作者机构:
湖南省疾病预防控制中心
作者简介:
基金信息:
DOI:
CLC:
扫 描 看 全 文
Determination of Benzo(a)pyrene and Bisphenol A in Drinking Water by Solid Phase Extraction/High Performance Liquid Chromatography. [J]. 34(7):848-851(2015)
DOI:
Determination of Benzo(a)pyrene and Bisphenol A in Drinking Water by Solid Phase Extraction/High Performance Liquid Chromatography. [J]. 34(7):848-851(2015)DOI:
Determination of Benzo(a)pyrene and Bisphenol A in Drinking Water by Solid Phase Extraction/High Performance Liquid Chromatography
A method was developed for the simultaneous determination of benzo(а)pyrene(Bap)and bisphenol A(BPA)in drinking water by high performance liquid chromatography(HPLC) with solid phase extraction(SPE).Benzo(a)pyrene and bisphenol A were concentrated from drinking water sample with an ENVI-18 column.The chromatographic separation was performed on a BEH-C18 column using acetonitrile - water as mobile phase with gradient elution.The benzo(a)pyrene and bisphenol A were detected with fluorescence detector(FLD).The results showed that the linear ranges of two compounds ranged from 0.1 to 20.0 μg/L with correlation coefficients(r)more than 0.999.High recoveries(86.1%-101%) and good reproducibility(2.9%-4.6%) were obtained.The limits of detection were 0.2 ng/L and 2.0 ng/L,respectively.The method was simple,feasible and accurate,and could be applied in the quantification of benzo(a)pyrene and bisphenol A in drinking water.
关键词
饮用水荧光检测器高效液相色谱固相萃取苯并(a)芘双酚A
Keywords
drinking waterfluorescence detector(FLD)high performance liquid chromatography(HPLC)solid phase extractionbenzo(а)pyrenebisphenol A
Determination of Haloacetic Acids in Drinking Water Using Ultra Performance Liquid Chromatography Coupled with Solid phase Extraction
Determination of Thiencarbazone-methyl and Its Metabolites in Different Environmental Matrixes by Liquid Chromatography- Tandem Mass Spectrometry
Simultaneous Determination of 12 Organophosphate Esters in Soil by Solid Phase Extraction/High Performance Liquid Chromatography-Tandem Mass Spectrometry
Determination of 53 Semi-volatile Organic Compounds in Solid Waste by Gas Chromatography-Mass Spectrometry with Microwave Extraction and Solid Phase Extraction Column Purification
Determination of New Red in Beverages and Prepared Wines by HPLC-MS/MS with Solid Phase Extraction Based on Metal Organic Frameworks UiO-66-NH2
Related Author
No data
Related Institution
国家地质实验测试中心
Key Laboratory of Pesticide Environmental Assessment and Pollution Control,Ministry of Environmental Protection,Nanjing Institute of Environmental Sciences,Ministry of Ecology and Environment of the People’s Republic of China
College of Carbon Neutral Modern Industry,College of Environmental and Resource Science,Fujian Normal University
Fujian Key Laboratory of Pollution Control and Resource Reuse
Fujian College Association Instrument Analysis Center,Fuzhou University